Add to Cart
Automated filter tester with high accuracy
Product Detail
Test flow | 30~95L/min |
Test area | 100cm2 |
Flow rate accuracy | ±3% |
Resistance accuracy | 1Pa |
Test repeatability of residual | ≤±10% |
Power supply | AC220V, 50 or 60 Hz |
Dimension | 750mm*700mm*1500mm |
Efficiency range | 10%~99.95% and 95%~99.999% are optional |
Aerosol | NaCl, DEHS, Paraffin oil, etc. |
Resistance range | 0~125Pa and 0~500Pa are optional |
Product Keywords | en149 automated filter tester,china cheap en149 automated filter tester,automated filter tester with high accuracy |
Product description
SC-FT-1806D Automatic Filter(Mask) Tester (Photometer) is designed to test the initial filtration efficiency, minimum filtration efficiency and airflow resistance of filter materials and respirator (mask). This instrument can comply with NIOSH 42 CFR 84, EN 149 and South Korean standards for health masks.
For filtration efficiency test, this machine can do single-machine for dual-aerosol (both oily and salty particles). Customers can choose double generator or single generator according to their own needs. Additional clean air source system is required if you choose the double-generator.
This instrument has the following characteristics:
1.We can offered diffierent language interface according to the customers' requirement.
2.Short testing time, a single test can be finished in 1 minute.
3.Controlled by microcomputer and operated by touch screen.
4.Data can be stored.
5.Installed ribbon printer, test results can be printed directly, and saved for a long time.
6.This tester can be connected to a computer by choosing the data acquisition software.
7.Stable performance, good data reproducibility.
8.Maintenance-free, suitable for production companies in long-term continuous use of quality control.
Technical Parameters
1. Flow range: adjutable 10~100 L / min, accuracy 1%
2. Filtration efficiency: (0.001 ~ 99.999)%
3. Resistance: 0 ~ 1000Pa, accuracy 2%
4. Test area: 100cm²
5. Aerosol: sodium chloride NaCl (optional DOP, DEHS, paraffin oil, corn oil)
6. Aerosol concentration: no more than 30mg / m³ (salinity), no more than 30mg / m³ (oily)
7. Test mode: initial efficiency and loading efficiency
8. Test temperature: (25 ± 5) ℃
9. Relative humidity: (30 ± 10)%
10. Dimensions: 780 × 880 × 1650mm
11. Weight: 190kg
12. Power source: AC220V 50Hz 300W
13. Air source: 0.4Mpa
14. Power supply: AC220V 50Hz
Scope of application
It is used to detect the filtration efficiency of particulates such as masks and filter materials. The system uses precise instruments to control the flow rate and aerosol concentration. It is easy to operate and has good stability.